Hioki IM7580A Impedance Analyser

Hioki IM7580A Impedance Analyser

Hioki LCR Meters and Impedance Analyzers range from 1mHz to 3GHz devices to suit a wide range of applications in the testing of electronic components.

The Hioki IM7580A impedance analyzer offers a top measurement time of 0.5ms over the complete frequency range, and both LCR and Analyzer Modes to meet applications for both high volume production inspections and research and development.

Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)

Measurement modes LCR mode, Analyzer mode (Sweeps with measurement frequency and measurement level), Continuous measurement mode
Measurement parameters Z, Y, θ, Rs (ESR), Rp, X, G, B, Cs, Cp, Ls, Lp, D (tanδ), Q
Measurable range 100 mΩ to 5 kΩ
Display range Z: 0.00 m to 9.99999 GΩ / Rs, Rp, X: ± (0.00 m to 9.99999 GΩ)
Ls, Lp: ± (0.00000 n to 9.99999 GH) / Q: ± (0.00 to 9999.99)
θ: ± (0.000° to 180.000°), Cs, Cp: ± (0.00000 p to 9.99999 GF)
D: ± (0.00000 to 9.99999), Y: (0.000 n to 9.99999 GS)
G, B: ± (0.000 n to 9.99999 GS), Δ%: ± (0.000 % to 999.999 %)
Basic accuracy Z: ±0.72 % rdg. θ: ±0.41°
Measurement frequency 1.0000 MHz to 300.00 MHz (5 digits resolution)
Measurement signal level Power: -40.0 dBm up to +7.0 dBm
Voltage: 4 mV to 1001 mVrms
Current: 0.09 mA to 20.02 mArms
Output impedance 50 Ω
Display 8.4-inch color TFT with touch screen
Measurement speeds *1 FAST: 0.5 ms (Analog measurement time, typical value)
Functions Contact check, Comparator, BIN measurement (classification), Panel loading/saving, Memory function, Equivalent circuit analysis, Correlation compensation
Interfaces EXT I/O (Handler), USB communication, USB memory, LAN, RS-232C (optional), GP-IB (optional)
Power supply 100 to 240 V AC, 50/60 Hz, 70 VA max.
Dimensions and mass Main unit: 215 mm (8.46 in) W × 200 mm (7.87 in) H × 268 mm (10.55 in) D, 6.5 kg (229.3 oz)
Test head: 61 mm (2.40 in) W × 55 mm (2.17 in) H × 24 mm (0.94 in) D, 175 g (6.2 oz)
Accessories Test head ×1, Connection cable ×1, Instruction manual ×1, LCR application disc (Communications user manual) ×1, Power cord ×1

Category:

Key Features

  • 1 MHz to 300 MHz testing source frequency
  • Fastest test speed of 0.5 msec
  • ±0.72% rdg. basic accuracy
  • Half-rack size body and palm-sized test head
  • Comprehensive contact check (via DCR testing, Hi-Z reject or waveform judgment)
  • Make frequency sweeps, level sweeps and time interval measurements in Analyzer Mode

Application Notes