Hioki DT4282 DIGITAL MULTIMETER
Hioki DT4282 DIGITAL MULTIMETER
£455.84 exc. VAT
Out of stock
Introducing the world-renowned DIGITAL MULTIMETER DT4282, the premier digital multimeter for laboratory and R&D applications. With superior accuracy and high response, this powerful device offers a wide range of measuring capabilities, including DC and AC current, conductance, and more.
The DT4282 features a low-pass filter function that eliminates harmonic components, allowing for accurate measurement of fundamental waves. It also offers peak value measurement and DC + AC voltage measurement for ripple voltage measurement in DC charging systems.
This digital multimeter also has a percentage display for 4 to 20 mA instrumentation signals, allowing for the conversion and display of instrumentation signals from transducers or other devices.
Designed and manufactured in Japan, the DT4282 is made with the highest quality materials and is backed by a 3-year warranty for peace of mind. Safety is paramount, and the DT4282 includes safety terminal shutters and optical communications to ensure electrical hazards are minimized.
With its advanced technological capabilities and unmatched precision, the DIGITAL MULTIMETER DT4282 is the ultimate tool for accurate and reliable measurements in any laboratory or R&D setting.
- Hioki DT4221: This model is a compact digital multimeter with a variety of measurement capabilities, including voltage, current, resistance, frequency, and temperature.
- Hioki DT4222: This model is similar to the DT4221 but without the capacitance measurement capability.
- Hioki DT4223: This model is similar to the DT4222 but with the capacitance and temperature measurement capabilities.
- Hioki DT4252: This model is designed for measuring low-level currents, with a range of 2mA to 20mA.
- Hioki DT4253: This model is similar to the DT4252 but with a higher current measurement range of 2mA to 200mA.
- Hioki DT4255: This model is designed for measuring high voltages and currents, up to 600V and 10A, respectively.
- Hioki DT4256: This model is similar to the DT4255 but with slightly higher voltage and current measurement capabilities (up to 1000V and 20A, respectively).
- Hioki DT4261: This model has a high resolution and accuracy and is designed for use in industrial settings.
- Hioki DT4281: This model is designed for high-speed measurement and can measure up to 120,000 times per second.
- Hioki DT4282: This model is similar to the DT4281 but also includes a built-in data logger for recording measurements over time.
Basic specifications (Accuracy guaranteed for 3 years, Post-adjustment accuracy guaranteed for 3 years)
DC Voltage range | 60.0 mV to 1000.0 V, 6 ranges, Basic accuracy: ±0.025 % rdg. ±2 dgt. |
---|---|
AC Voltage range* | 60.00 mV to 1000.0 V, 6 ranges, Frequency characteristics: 20 Hz – 100 kHz Basic accuracy 40 – 65 Hz : ±0.2 % rdg. ±25 dgt. (True RMS, crest factor 3) |
AC + DC Voltage range* | 60.00 mV to 1000.0 V, 6 ranges, Frequency characteristics: 20 Hz – 100 kHz Basic accuracy 40 – 65 Hz : ±0.3 % rdg. ±25 dgt. (True RMS, crest factor 3) |
Resistance range | 60.000 Ω to 600.0 MΩ, 8 ranges, Basic accuracy: ±0.03 % rdg ±2 dgt |
DC Current range | 600.00 μA to 10.000 A, 6 ranges. Basic accuracy: ±0.05 % rdg ±5 dgt |
AC Current* range | 600.00 μA to 10.000 A, 6 ranges. Basic accuracy 45 – 65 Hz : ±0.6 % rdg ±5 dgt (True RMS, crest factor 3). |
Peak DC | DC V measurement: Signal width 4 msec or more (single), 1 msec or more (repeated). |
Peak AC | AC V, DC/AC A measurement: Signal width 1 msec or more (single), 250 μsec or more (repeated) |
Capacitance range | 1.000 nF to 100.0 mF, 9 ranges, Basic accuracy: ±1.0 % rdg ±5 dgt |
Frequency range | AC V, DC+AC V, AC A measurement,99.999 Hz to 500.00 kHz, 5 ranges, ±0.005 % rdg ±3 dgt |
Continuity check | Continuity threshold: 20/50/100/500 Ω, Response time: 10 ms or more |
Diode test | Open terminal voltage: 4.5 V or less, Testing current 1.2 mA or less, Threshold of forward voltage: 0.15 V to 3 V, seven stages |
dB conversion | Standard impedance setting (dBm), 4 Ω to 1200 Ω, 20 stages Display dB conversion value of AC voltage (dBV) |
Temperature (thermocouples) |
K: -40.0 °C to 800.0 °C (-40.0 °F to 1472.0 °F) Add accuracy of the Thermocouple probe to main unit accuracy: ±0.5 % rdg ±3 °C |
Other functions | Filter function, Display value hold, Auto hold, Max/Min value display, Sampling select, Relative display, Measurement memory (400 data), Auto-power save |
Display | Main and Sub displays: 5-digits LCD, max. 6000 digits |
Display refresh rates | 5 times/s (Capacitance measurement: 0.05 to 2 times/s, depending on measured value, Temperature: 1 time/s ) |
Power supply | LR6 (AA) alkaline batteries ×4, Continuous use: 100 hours |
Dimensions and mass | 93 mm (3.66 in)W × 197 mm (7.76 in)H× 53 mm (2.09 in)D, 650 g (22.9 oz) (including test leads holder and batteries) |
Accessories | Test lead L9207-10 ×1, Instruction manual ×1, LR6 alkaline battery ×4 |
* Zero-suppression: For small inputs below the guarantee range, zero is effectively displayed