HiokiCT6862-05 AC/DC Current Sensor

HiokiCT6862-05 AC/DC Current Sensor

Hioki pass-through AC/DC current sensors are best-in-class devices for use with power meters, Memory HiCorders and high performance oscilloscopes. The Hioki CT6862-05 is a DC to 1 MHz/50A sensor, and this is ideal for capturing current signals from switching power supplies, inverters, power conditioners, etc.

CAT III 1000 V

Use with an instrument with a high input impedance of 1 MΩ or higher.

Replacing the discontinued Hioki CT6862 50 A AC/DC, PL23 terminal

Basic specifications

Frequency characteristicsAmplitude: DC to 10 MHz
Phase: DC to 1 MHzOutput voltage40 mV/A rated (This device outputs AC+DC voltage via the Sensor Unit)

Rated current 50 A AC/DC
Maximum input Within the derating range, up to ±150 A peak (design value) allowed at 40°C or less for 20 ms or less
Accuracy

DC (±0.03% rdg. ±0.002% f.s., no phase specification)

45 Hz ≤ f ≤ 66 Hz (±0.03% rdg. ±0.007% f.s., ±0.05°)

Specified up to 1 MHz

Max. rated voltage to earth 1000 V CAT III
Core diameter φ 24 mm (0.94 in)
Operating temperature, humidity -40°C to +85°C, 80% RH or less (with no condensation)
Power supply Power suppled via the Power Analyzer PW8001, PW6001, PW3390, Sensor Unit CT9555, CT9556, CT9557, or 3CH CURRENT UNIT U8977
Power consumption 4 VA max. (at 50 A/55 Hz, ±12 V power requirement)
Dimensions and mass 70mm W × 100mm H × 53mm D, 340g, cord length: 3m
Accessories Instruction manual ×1, Mark bands ×6, Operating Precautions

Compatible models…CT6862-05

CT6872 OK
Power Analyzer PW8001, PW6001, PW3390 OK
Power Analyzer 3390 Requires CT9901
Power HiTester 3193 series Requires CT9901
Current Unit 8971 Requires the 9318, CT9901
3CH Current Unit U8977 OK

Category:

Key Features

  • Combined accuracy with HIOKI power analyzer PW8001, PW6001 and PW3390 is specified. For details of combined accuracy, refer to the instruction manual
  • Wide-bandwidth DC to 10 MHz excellent frequency characteristics
  • Applications in the fields of electric and hybrid electric vehicles
  • Wide operating temperature range(-40°C to 85°C) fit for automobile applications
  • Ideal for evaluation of solar power generation and fuel cells to measure battery charge and discharge and the secondary side of inverters
  • For observing waveforms to be used with the oscilloscopes or Memory HiCorders (use with SENSOR UNIT)

Application Notes