HiokiCT6862-05 AC/DC Current Sensor
Hioki pass-through AC/DC current sensors are best-in-class devices for use with power meters, Memory HiCorders and high performance oscilloscopes. The Hioki CT6862-05 is a DC to 1 MHz/50A sensor, and this is ideal for capturing current signals from switching power supplies, inverters, power conditioners, etc.
CAT III 1000 V
Use with an instrument with a high input impedance of 1 MΩ or higher.
Replacing the discontinued Hioki CT6862 50 A AC/DC, PL23 terminal
Basic specifications
Frequency characteristicsAmplitude: DC to 10 MHz
Phase: DC to 1 MHzOutput voltage40 mV/A rated (This device outputs AC+DC voltage via the Sensor Unit)
Rated current | 50 A AC/DC |
---|---|
Maximum input | Within the derating range, up to ±150 A peak (design value) allowed at 40°C or less for 20 ms or less |
Accuracy
DC (±0.03% rdg. ±0.002% f.s., no phase specification) 45 Hz ≤ f ≤ 66 Hz (±0.03% rdg. ±0.007% f.s., ±0.05°) Specified up to 1 MHz |
|
Max. rated voltage to earth | 1000 V CAT III |
Core diameter | φ 24 mm (0.94 in) |
Operating temperature, humidity | -40°C to +85°C, 80% RH or less (with no condensation) |
Power supply | Power suppled via the Power Analyzer PW8001, PW6001, PW3390, Sensor Unit CT9555, CT9556, CT9557, or 3CH CURRENT UNIT U8977 |
Power consumption | 4 VA max. (at 50 A/55 Hz, ±12 V power requirement) |
Dimensions and mass | 70mm W × 100mm H × 53mm D, 340g, cord length: 3m |
Accessories | Instruction manual ×1, Mark bands ×6, Operating Precautions |
Compatible models…CT6862-05
CT6872 | OK |
Power Analyzer PW8001, PW6001, PW3390 | OK |
Power Analyzer 3390 | Requires CT9901 |
Power HiTester 3193 series | Requires CT9901 |
Current Unit 8971 | Requires the 9318, CT9901 |
3CH Current Unit U8977 | OK |