PMM Narda 9010/Click4E Click Meter CISPR 14-1 – Discontinued
The PMM Narda Click4E Click Meter when used with the PMM 9010F FFT RF receiver with relevant software key provides a four channel Click Analyser fully compliant to CISPR 14-1 and CISPR 16-1-2.
Smart Measure: how to save test time
Depending on the EUT, the discontinuous disturbances to be assessed as “clicks” may saturate the conventional, analog type Click Analyzers, thus giving false results and repeat testing. The Smart Measure function of the 9010F Click option immediately detects and stores any overloading that may occur during the 1st phase (determination of the Click Rate N) and then can stop and warn the operator or continue the test; the clicks recorder in overloading condition are then recorded in the test report. If no overloading occurred, the stored values can be used and the 2nd measurement phase can be avoided, reducing the test time by half.
The 9010/Click4E Metre Option gives the possibility to measure simultaneously up to four frequencies (channels) by the way reducing the time needed to complete the test.