Hioki 8972 DC/RMS Unit

Hioki 8972 DC/RMS Unit

Introducing the DC/RMS UNIT 8972, a versatile and high-resolution waveform measurement device designed for precision and accuracy in a wide range of applications. This advanced unit is equipped with a range of features that make it a must-have tool for engineers, technicians, and researchers.

The DC/RMS UNIT 8972 is designed to measure both instantaneous and RMS waveforms with high 12-bit A/D resolution, providing you with accurate and reliable readings every time. With its isolated 2-channel input, you can measure signals from two different sources simultaneously, saving you time and effort.

The DC/RMS UNIT 8972 is also built to withstand high voltages, with a maximum rated voltage to ground of 300 V AC, DC (CAT II). This makes it suitable for use in a range of industrial and laboratory settings, where high voltages are often present.

One of the standout features of the DC/RMS UNIT 8972 is its wide bandwidth RMS rectifier, which can measure signals from low to high frequency, from DC to 30 Hz to 100 kHz. This allows you to measure both AC and DC signals accurately, making it ideal for a range of applications.

Finally, with the DC/RMS UNIT 8972, you can monitor instantaneous waveforms without the need to pass through the rectifying circuit, providing you with even more flexibility and accuracy when measuring signals.

Overall, the DC/RMS UNIT 8972 is an essential tool for anyone working with waveform measurements. Its high resolution, isolated input, and wide bandwidth RMS rectifier make it an indispensable addition to any laboratory or industrial setting.

Measurement functions Number of channels: 2, for voltage measurement, DC/RMS selectable
Input connectors Isolated BNC connector (input impedance 1 MΩ, input capacitance 30 pF),
Max. rated voltage to earth: 300 V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage)
Measurement range 5 mV to 20 V/div, 12 ranges, full scale: 20 div, (AC voltage for possible measurement/display using the memory function: 280 V rms)
Low-pass filter: 5/50/500 Hz, 5 k/100 kHz
Measurement resolution 1/100 of measurement range (using 12-bit A/D conversion and when installed in 8847)
Highest sampling rate 1 MS/s (simultaneous sampling across 2 channels)
Measurement accuracy ±0.5 % of full scale (with filter 5 Hz, zero position accuracy included)
RMS measurement RMS amplitude accuracy: ±1 % of full scale (DC, 30 Hz to 1 kHz), ±3 % of full scale (1 kHz to 100 kHz),
Response time: SLOW 5 s (rise time from 0 to 90% of full scale), MID 800 ms (rise time from 0 to 90% of full scale), FAST 100 ms (rise time from 0 to 90% of full scale), Crest factor: 2
Frequency characteristics DC to 400 kHz -3 dB, with AC coupling: 7 Hz to 400 kHz -3dB
Input coupling AC/DC/GND
Max. allowable input 400 V DC (the maximum voltage that can be applied across input pins without damage)
Dimensions and mass 106 mm (4.17 in) W × 19.8 mm (0.78 in) H × 196.5 mm (7.74 in) D, 250 g (8.8 oz)
Accessories None

Category:

Key Features

  • 12-bit A/D resolution
  • From low to high frequency, wide bandwidth RMS rectifier installed
  • (DC, or 30 Hz to 100 kHz)
  • Monitor instantaneous waveforms without the need to pass through the rectifying circuit
  • Isolated input, up to 300 V AC/DC from the earth

 

Application Notes

Manufacturer

Downloads

There is no technical specification to download at this time, if you would like more information regarding this product please click here