EMSense for Battery-Powered E-Field Probes Model 7007-001

EMSense for Battery-Powered E-Field Probes Model 7007-001

ETS-LINDGREN’S EMSense MODULAR PLUG-IN CARD provides an interface for ETS-Lindgren’s battery-powered E-field probes.

Each EMSense modular plug-in card supports one ETS-Lindgren E-field probe. EMSense and ETS-Lindgren probes are fully supported by TILETM and other test automation software packages.

EMSense is fully compatible with all ETS-Lindgren E-field battery-powered probes, covering a frequency range of 10 kHz – 40 GHz and a dynamic range of 0.5 – 800 V/m and occupies one slot in the EMCenter platform, allowing additional slots to be occupied by modular plug-in cards for positioning controllers, multiple RF switches and more.

Performance Specifications (E-Field Probes)

MODEL FEATURE

HI-6022

HI-6005 

HI-6053

Detection Isotropic

Isotropic (X, Y and Z Axis Readings)

Isotropic (X, Y and Z Axis Readings)

Isotropic (X, Y and Z Axis Readings)

Calibrated Frequency Range

10 kHz to 1 GHz

100 kHz to 6 GHz

10 MHz to 40 GHz

Frequency Response (Typical)

10 kHz to 30 kHz +0.5, – 2.5 dB
30 kHz to 1 GHz ± 1.0 dB

 500 kHz to 2 GHz +1.0, -2.5 dB
2 to 5 GHz +3.5, -4.0 dB
5 to 6 GHz +1.0, -6.0 dB

10 to 100 MHz + 3.0, -4.0 dB
100 MHz to 1 GHz +3.0, -0.50 dB
1 to 18 GHz +4.0, – 2.0 dB
18 to 40 GHz +3.0, -4.5 dB

Frequency Response with Correction

± 0.9 dB

± 0.9 dB

10 MHz to 18 GHz ± 0.9 dB
18 to 40 GHz ± 1.1 dB

Dynamic Range

2.0 to 800 V/m (Single Range)

0.5 – 800 V/m (Single Range)

2.0 to 800 V/m (Single Range)

Resolution

 0.01 V/m

 0.01 V/m

 0.01 V/m

Isotropic Deviation

 ± 0.5 dB @ 400 MHz

 ± 0.5 dB @ 400 MHz

 ±1.0 dB < 18 GHz

Linearity

± 0.5 dB @ 27 MHz

± 0.5 dB @ 27 MHz

±0.5 dB @ 1 GHz

Sample Rate

60 Samples/Second Maximum

60 Samples/Second Maximum

60 Samples/Second Maximum

Overload Withstand

>1500 V/m CW

>1500 V/m CW

>1500 V/m CW

Category: