Hioki IR4056 INSULATION TESTER

Hioki IR4056 INSULATION TESTER

Hioki insulation testers or megohmmeters feature all-in-one designs including built-in rugged cases and test lead storage compartments. The Hioki IR4056 is a 5-range, 50V to 1000V digital insulation resistance tester with efficient functions and features to help shorten work time associated with field insulation testing. Other models available, IR4057 (High Speed) and IR4058 (High Speed with Bluetooth)

 

 

Basic specifications (Accuracy guaranteed for 3 years, Post-adjustment accuracy guaranteed for 3 years)

Rated output voltage 50 V DC 125 V DC 250 V DC 500 V DC 1000 V DC
Effective maximum indicated value 100 MΩ 250 MΩ 500 MΩ 2000 MΩ 4000 MΩ
Accuracy 1st effective measuring range MΩ ±2 % rdg. ±2 dgt.
0.200 – 10.00
±2 % rdg. ±2 dgt.
0.200 – 25.0
±2 % rdg. ±2 dgt.
0.200 – 50.0
±2 % rdg. ±2 dgt.
0.200 – 500
±2 % rdg. ±2 dgt.
0.200 – 1000
Lower limit resistance 0.05 MΩ 0.125 MΩ 0.25 MΩ 0.5 MΩ 1 MΩ
Overload protection 600 V AC (10s) 660 V AC (10s)
DC voltage range 4.2 V (0.001 V resolution) to 600 V (1 V resolution), 4 ranges,
Accuracy: ±1.3 % rdg. ±4 dgt., Input resistance: 100 kΩ or higher
AC voltage range 420 V (0.1 V resolution) / 600 V (1 V resolution), 2 ranges, 50/60 Hz,
Accuracy: ±2.3% rdg. ±8 dgt., Input resistance: 100 kΩ or higher, Average rectifier
Low resistance range For checking the continuity of ground wiring, 10 Ω (0.01 Ω resolution) to 1000 Ω (1 Ω resolution), 3 ranges, Basic accuracy: ±3 % rdg. ±2 dgt., testing current 200 mA or more (at 6 Ω or less)
Display Semi-transmissive FSTN LCD with back lighting, bar-graph indicator
Response time Approx. 0.8 second for PASS/FAIL decision (based on in-house testing)
Other functions Live circuit indicator, Automatic electric discharge, Automatic DC/AC detection, Comparator, Drop proof, Auto power save
Power supply LR6 (AA) alkaline batteries ×4, Continuous use: 20 hours (Comparator off, backlight off, 500 V range, no load)
Number of measurements: 1000 times (at 5 s ON, 25 s OFF cycle, insulation measurement of lower limit resistance value to maintain nominal output voltage)
Dimensions and mass 159 mm (6.26 in) W × 177 mm (6.97 in) H× 53 mm (2.09 in) D, 600 g (21.2 oz) (including batteries, excluding test leads)
Accessories [IR4056-20] Test lead L9787 ×1, Neck strap ×1, Instruction manual ×1, LR6 (AA) alkaline batteries ×4
[IR4056-21] Test lead set with remote switch L9788-11 ×1, Neck strap ×1, Instruction manual ×1, LR6 (AA) alkaline batteries ×4

Category:

Key Features

• 5-range testing voltage of 50 V/100 MΩ to 1000 V/4000 MΩ

• Stable & medium-speed digital readings, 0.8 second response time of PASS/ FAIL decisions

• Drop proof onto concrete from 1m (3.28 feet)

• Bright LED, luminous LCD, test lead with bright LED lamp to illuminate near hand (Also available in the IR4056-21)

• Continuity check via 200 mA testing

• Built in AC/DC voltage meter, useful for testing solar power generation systems and electric vehicles

 

Application Notes