Hioki 3930 HIGH VOLTAGE SCANNER

Hioki 3930 HIGH VOLTAGE SCANNER

The Hioki 3930 is an essential accessory for your Hioki 3153 automatic insulation/withstanding tester or sequencer. This passive auxiliary tester is designed to make the testing process more efficient, safe, and hassle-free. With the Hioki 3930, you can test multiple points with minimal connection changes.

One of the key features of the Hioki 3930 is the output of the input high voltage from a user-selected channel. This means that you can choose the channel you want to test and apply high voltage to it. The Hioki 3930 also comes with eight channels per unit (single mode) and up to 32 channels (four connected units). This allows you to test multiple channels at once, making the testing process faster and more efficient.

The Hioki 3930 also comes with isolated high-voltage I/O, control signal lines, and power supply. This feature ensures that your testing process is safe and reliable. The isolation prevents any interference or crosstalk between the channels, ensuring accurate and reliable results.

You can control the Hioki 3930 using the 3153 program function or with a standard sequencer. This makes the testing process more flexible and allows you to choose the control method that works best for you.

In summary, the Hioki 3930 is an essential accessory for your Hioki 3153 automatic insulation/withstanding tester or sequencer. With its key features such as the output of the input high voltage, multiple channels per unit, isolated high-voltage I/O, and flexible control options, the Hioki 3930 makes the testing process faster, safer, and more efficient.

Basic specifications (Accuracy guaranteed for 3 years, Post-adjustment accuracy guaranteed for 3 years)

Operation modes Multi-mode: Scanning of user-selected points for high 4 ch / low 4 ch
Single mode: Common scan of high 8 ch – common
Rated voltage used 5 kV AC / 5 kV DC
Operation indications Lamps light up when power is supplied and when a specified channel is operating

[Relay area]

Max. open and closed voltage 5000 V DC, 5000 V AC
Max. open and closed current 1.0 A (open and closed capacity: 50 W)
Contact point indirect contact resistance 500 mΩ or less, with 1 mA AC
Contact point max. capacity 50 W
Time Operation time: 6 ms or less, Recovery time: 6 ms or less
Power supply VSCV 24 V DC, ±10% (applied using the control signal input connector), 12 VA max.
Dimensions and mass 316 mm (12.44 in) W × 100 mm (3.94 in) H × 350 mm (13.78 in) D, 4.2 kg (148.1 oz)
Accessories Control input connector connection cable ×1, H.V. Test Lead 9615-01 (red) ×8, H.V. Test lead (black) ×1, Grounding cable ×1, Instruction manual ×1

Category:

You may also like…

Key Features

• Output of the input high voltage from a user-selected channel

• 8 ch per unit (single mode), with up to 32 ch (4 connected units)

• Isolated high-voltage I/O, control signal lines, and power supply

• Control using the 3153 program function or with a standard sequencer

Application Notes