Hioki RM3542 Resistance HiTester
Hioki DC resistance meters measure a broad range of resistance values at a high level of precision.
The Hioki RM3542 deliver high speed, high precision, and high reliability, indispensable for the shipping inspection of devices such as resistors and EMI components. The resistance meter also supports chip resistor sampling inspections, helping resistor manufacturers improve productivity.
Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Resistance range | [at Low Power OFF] 100 mΩ range (Max. 120.0000 mΩ, 0.1 μΩ resolution) to 100 MΩ range (Max. 120.0000 MΩ, 100 Ω resolution), 10 steps [at Low Power ON] 1000 mΩ range (Max. 1200.000 mΩ, 1 μΩ resolution) to 1000 Ω range (Max. 1200.000 Ω, 1 mΩ resolution), 4 steps |
---|---|
Display | Monochrome graphic LCD 240 × 64 dot, white LED backlight |
Measurement accuracy | [with SLOW mode, at 100 mΩ range] ±0.015 % rdg. ±0.002 % f.s. [with SLOW mode, at 1000 Ω range] ±0.006 % rdg. ±0.001 % f.s. (the best case) |
Testing current | [at 100 mΩ range] 100 mA DC to [at 100 MΩ range] 100 nA DC |
Open-terminal voltage | 20 V DC max. |
Sampling rate | FAST, MEDIUM, SLOW, 3 settings |
Measurement times | [at 100 Ω /1000 Ω ranges, with Low Power OFF] FAST: 0.9 ms, MED: 3.6 ms, SLOW: 17 ms (minimum time) |
Integration time | 0.1 ms to 100.0 ms, or 1 to 5 PLC at 50 Hz, 1 to 6 PLC at 60 Hz Note: PLC = one power line cycle (mains wave-form period) |
Other functions | Comparator (compare setting value with measurement value), Delay (set to allow for mechanical delay of trigger input and probing, or set to allow for measurement object response), OVC (offset voltage compensation), Measurement fault detection, Probe short-circuit detection, Improve contact, Auto-memory, Statistical calculations, Settings monitor (when using two instruments, a difference in settings causes warning notification), Retry, Trigger function. etc,. |
Interfaces | RS-232C, Printer, GP-IB (Model RM3542-01) |
External I/O | Trigger, Hold input, Comparator output, Settings monitor terminal |
Power supply | 100 to 240 V AC, 50/60 Hz, 30 VA max. |
Dimensions and mass | 260 mm (10.24 in) W × 88 mm (3.46 in) H × 300 mm (11.81 in) D, 2.9 kg (102.3 oz) |
Accessories | Power cord ×1, EXT. I/O male connector ×1, Instruction manual ×1, Operation guide ×1 |
Model No. (Order Code)
RM3542 | |
---|---|
RM3542-01 | Built in GP-IB interface |