Hioki CT6843A AC/DC Current Clamp

Hioki CT6843A AC/DC Current Clamp

Hioki’s clamp type AC/DC current sensors can be connected directly to Hioki Power Analyzers and Memory HiCorders.
Also, it can be connected to Power Meters, Oscilloscopes, DMMs, etc. by using an optional sensor unit.

Use with an instrument with a high input impedance of 1 MΩ or higher.

Replacing the discontinued Hioki CT6861 200 A AC/DC, PL23 terminal and Hioki CT6861-05 200 A AC/DC, ME15W terminal

 

Basic specifications

Rated current 200 A AC/DC
Frequency bandwidth Amplitude: DC to 700 kHz
Core diameter φ 20 mm
Max. allowable input ±600 A peak (Within 20 ms in an environment of 40°C or less)
Output voltage 10 mV/A
Output resistance 50 Ω ± 10 Ω/td>
Amplitude accuracy DC: 0.2% + 0.02%, DC < f ≤ 100 Hz: 0.2% + 0.01%
Power consumption 6 VA max. (at 200 A/55 Hz, ±12 V power requirement)
Linearity ±20 ppm Typical
Common-Mode Voltage Rejection Ratio (CMRR) DC to 1 kHz:150 dB or greater

1 kHz to 10kHz:135 dB or greater

10 kHz to 100 kHz:115 dB or greater

100 kHz to 500 kHz:95 dB or greater

(effect on output voltage and common mode voltage)

Automatic phase correction Automatically performs phase correction when connected to PW8001
Operating temperature, humidity -40 °C to +85 °C, 80% RH or less
Withstand voltage AC 4260 V
Standards Safety: IEC 61010 Type D, EMC: IEC 61326
Power supply Power suppled via the Power Analyzer PW8001, PW6001, PW3390, Sensor Unit CT9555, CT9556, CT9557, or 3CH CURRENT UNIT U8977
Dimensions and mass 153mm W × 67mm H × 25mm D, 370g, cord length: 3 m
Accessories Instruction manual ×1, Mark bands ×6, Carrying Case ×1

Category:

You may also like…

Key Features

  • Frequency bandwidth DC to 700 kHz
  • Rated current 200 A AC/DC
  • Power supplied via the measurement instrument (when connecting Hioki Power Analyzer or Memory HiCorder)
  • Ideal for use in environmental testing with broad -40°C to 85°C temperature range
  • Ideal for EV inverter evaluation and PV power generation PCS evaluation
  • Single-handed operation and robust locking mechanism